LIGO Document T2000182-x0
- Collection of useful DCC documents and links regarding Working Standard responsivity ratio measurements performed in the Pcal lab.
The pcal_svn is located at: _______________.
Lab responsivity ratio measurements are located in pcal_svn/Photon_Calibrator/LabData. The directories located here (ex: WSH_GS) indicate the two standards used in the measurements, with the first/second standard being the standard in the numerator/denominator of the responsivity ratio.
Measurements listed in these directories may have "t", "x", or "i" as the first letter.
"t": temporary directory, reserved for measurements that were taken recently but have not been fully vetted for data quality.
"x": unused directory, taken out of the measurement pool due to bad quality/changes to the working standard/other reasons. Should ideally have reason for removal in the notes.txt file.
"i": experiment directory, not to be used for responsivity estimates, but useful for testing the state of our system (ex: temperature dependence measurements).
"D" as the first letter means the measurement has been vetted and approved.
Procedure for in-lab responsivity ratio measurements: T1400442
Working Standard configuration/setup: T070210
Most recent changes to the standards: T1800359
Responsivity ratio deliberations for beginning of O3: T1900134
Running Pcal analysis code: T1900758
The general directory for end station calibration measurements is located at: T2000191
DCC Version 3.5.0, contact
DCC Help